Buy or Rent Buy or Rent Heron dipperLog 32 Groundwater Data Logger

The most economical data logger in our lineup, the dipperLog 32 has a maximum storage capcity of 32,000 data points and expected battery life up to 5 years.

The dipperLog 32, only available in a 30m/100ft pressure range, is a reliable solution for short term monitoring of groundwater levels, wetland studies, pump and slug testing.

All Heron loggers are available in various deployment options:

Suspended Version – pull the dipperLog out of the well to download

Direct Read Version – enables you to do real time readings and download your data while the dipperLog is still in the well.

Reel Version – a portable solution, ideal for short term slug and pump testing


The dedicated, intuitive software makes the dipperLog quick and easy to program and launch. Simply recover your data, save your findings in a tabular display and graph, view both parameters side by side in a split screen.

  • Buy
Clear
SKU: 5401-4-Ti-1 Categories: ,
  • Memory – 32,000 data sets (depth and temperature of water)
  • Download Speed – 32,000 data sets in 90 seconds
  • Battery – up to 5 years (depending on usage)
  • Calibration Certificate provided with each logger
  • Machined, 316 marine grade stainless steel body and transducer
  • Able to convert deployment options easily in the field
Features  
# of Data Points 32,000
Download Speed 21,333.33/min
Pressure Ranges 30m(100ft)
Overpressure Rating 1.5X
Operating Temperature -10°C to 50°C
Expected Battery Life Up to 5 Years (depending on usage)
Warranty 1 Year
   
Temperature Specifications  
Accuracy +/-0.5°C (-25°C to +85°C)
Resolution 0.0625°C
   
Pressure Specifications  
Accuracy Typical 0.05%FS
Max 0.1%FS
Resolution 0.006%FS
   
Physical Characteristics  
Wetted Material 316 Stainless Steel & Delrin®
O-Rings Rubber Buna
Transducer Material 316 Stainless Steel Piezoresistive Silicon
Length 105mm
Diameter 22mm
Weight 130g
   
Deployment Options  
Suspended  
Direct Read  
Reel  
   
Note: In “Log Time” mode the dipperLog increases the readings from 1 second by an additional 1 second for 255 readings. The final reading interval is 4 min. and 14 sec. and the logger then stops logging. This mode logs for a total of 9 hours before stopping.
   
Barometric Compensation The first reading can be used for all readings unless a barLog is used. Automatic compensation when using a designated barLog. Manual compensation in Excel.
   
Direct Read Cable 1/8″ Polyurethane

Product Manual (PDF)
Specifications (PDF)