GilAir Plus Personal Air Sampling Pump
The GilAir® Plus features wide dynamic flow and pressure ranges to address all personal air sampling methods with flow ranges between 20 and 5,000 cc/min. With patent pending QuadModeSM air sampling technology, the GilAir Plus can perform both high-flow constant pressure and constant flow (450-5,000 cc/min) and low-flow constant pressure and constant flow (20-499 cc/min) with a single pump without external adaptors. It’s like getting two pumps in one.
GilAir Plus is a smart air sampling pump that provides ambient or standard temperature and pressure corrected display (STP model) and data-logging for up to 16 sampling events (STP and Datalog model). The pump has high back pressure capability of up to 40″ H2O in high flow and up to 25″ H2O in low flow. An automatic self check system ensures sample accuracy. In the event of a fault condition, the pump’s selectable automatic fault recovery feature attempts to restart the pump up to 10 times every 3 minutes to determine if the fault condition has been cleared.
Other configurations and options available. Please contact your local Eco-Rental Solutions representative for assistance.
Additional Information
One Air Sampling Pump with All the Features You Need and Want
3 Prime Updated Technologies
Motion Detection
Bluetooth Connectivity
Gilian CONNECT Mobile App
Also Features:
QuadModeSM Air Sampling Technology
Selectable Automatic Fault Recovery
Standard Temperature & Pressure (STP) Correction
Superior Design & Comfort
Real-time Clock with Data-logging
Performance
Firmware
Current Version: 2.5.4 (6/20/17)
Flow Range
Constant flow 20 – 5,000 cc/min. without external adaptors
Constant pressure 1 – 5,000 cc/min without external adaptors
Flow Modes
Constant Flow or High and Low Flow Constant Pressure
Accuracy
Flow Display ± 5% or 3 cc/min of set flow, whichever is larger
Constant Flow Control ± 5% or 3 cc/min of set flow, whichever is larger
Constant Pressure Control ± 10% of back-pressure
Back Pressure Capability
(8-hour runtime) 5000 cc/min. up to 12” H20 back pressure
4000 cc/min. up to 20” H20 back pressure
3000 cc/min. up to 30” H20 back pressure
2000 cc/min. up to 30” H20 back pressure
1000 cc/min. up to 35” H20 back pressure
450 – 1000 cc/min. up to 40” H20 back pressure
20 – 449 cc/min. up to 40” H20 back pressure
Run Time 8 hour minimum
Flow Fault If flow changes exceed 5% within back pressure specifications, fault notification appears. If fault exceeds 30 seconds, pump shuts down. Selectable automatic fault recovery allows the pump to attempt restart every 3 minutes for up to 10 times or to hold until manual intervention.
Environmental
Temperature Ranges
Operating 32°F to 113°F (0°C to 45°C)
Storage -4°F to 113°F (-20°C to 45°C)
Charging 41°F to 104°F (5°C to 40°C)
Humidity Ranges
Operating 5-95 %RH, non-condensing
Storage 5-98 %RH, non-condensing
GENERAL
Display Flow rate, sample time, and sample volume in actual conditions
Controls
Indicators Green and Red LED’s
Dimensions 4.3 x 2.4 x 2.4 inches (10.9 x 6.1 x 6.1 centimeters)
Weight 20.5 ounce (581 grams)
Warranty 2 years from original shipment, 5 years for pump keypad, and 1 year on rechargeable NiMH battery pack
ELECTRICAL
Power Options Removable rechargeable nickel-metal hydride (NiMH) battery pack, optional alkaline battery pack, or DC power
Battery Level Indicator On-screen icon displays battery level
Interface Connectors All pump interface is carried out via the Dock or keypad
Dock Functions Battery charging, computer interface, and auto-calibration
Charging Time Under 3.5 hours