Rent YSI 6920 V2 Water Quality Sonde

Rent YSI 6920 V2 Water Quality Sonde

The YSI 6920 V2 sonde is an economical water quality logging system, ideal for long-term in situ monitoring and profiling.

Real-time turbidity monitoring, dissolved oxygen monitoring, algae monitoring, and more.

 

  • Field-replaceable sensors
  • Choice between one or two optical sensor ports
  • Self-cleaning optical sensors and anti-fouling components for extended deployment
  • RS-232 and SDI-12 communications
  • Battery power for long-term deployments

 

 

  • Source water monitoring
  • Dredging studies
  • Long-term unattended monitoiring
  • Spot sampling with complete sensor suite

 

Medium Fresh, sea or polluted water
Temperature Operating
Storage
-5 to +50°C
-10 to +60°C
Communications RS-232, SDI-12
Software EcoWatch®
Dimensions Diameter
Length
Weight
2.85 in, 7.24 cm
18 in, 45.7 cm
4 lbs, 1.8 kg
Power External
Internal
12 V DC
8 AA-size alkaline batteries
Certifications  

CE, EU Battery Compliance, FCC, IP-67, WEEE, and MCERTS
Assembled in the USA

 

  Range Resolution Accuracy
ROX™
Optical Dissolved Oxygen•
% Saturation
0 to 500% 0.1% 0 to 200%: ±1% of reading or 1% air saturation, whichever is greater;
200 to 500%: ±15% of reading
ROX™
Optical Dissolved Oxygen•
mg/L
0 to 50 mg/L 0.01 mg/L 0 to 20 mg/L: ± 0.1 mg/L or 1% of reading, whichever is greater;
20 to 50 mg/L: ±15% of reading
Dissolved Oxygen••
% Saturation
6562 Rapid Pulse™ Sensor*
0 to 500% 0.1% 0 to 200%: ±2% of reading or 2% air saturation, whichever is greater;
200 to 500%: ±6% of reading
Dissolved Oxygen••
mg/L
6562 Rapid Pulse™ Sensor*
0 to 50 mg/L 0.01 mg/L 0 to 20 mg/L: ± 0.2 mg/L or 2% of reading, whichever is greater;
20 to 50 mg/L: ±6% of reading
Conductivity•••
6560 Sensor*
0 to 100 mS/cm 0.001 to 0.1 mS/cm
(range dependent)
±0.5% of reading + 0.001 mS/cm
Salinity 0 to 70 ppt 0.01 ppt ±1% of reading or 0.1 ppt,
whichever is greater
Temperature
6560 Sensor*
-5 to +50°C 0.01°C ±0.15°C
pH
6561 Sensor*
0 to 14 units 0.01 unit ±0.2 unit
ORP -999 to +999 mV 0.1 mV ±20 mV
Depth Medium
Shallow
Vented Level
0 to 200 ft, 61 m
0 to 30 ft, 9.1 m
0 to 30 ft, 9.1 m
0.001 ft, 0.001 m
0.001 ft, 0.001 m
0.001 ft, 0.001 m
±0.4 ft, ±0.12 m
±0.06 ft, ±0.02 m
±0.01 ft, 0.003 m
Turbidity•
6136 Sensor*
0 to 1,000 NTU 0.1 NTU ±2% of reading or 0.3 NTU,
whichever is greater**

Product Manual (PDF)